Description
Drager Infinity Delta XL Features
The Drager Infinity Delta XL can monitor the vital signs of adult, pediatric, and neonatal patients with various acuity levels. Providing continuous monitoring and data collections, the Infinity Delta XL can be utilized at the bedside, when transporting, or anywhere in between. This patient monitor can connect to the Infinity Network via the Infinity Docking Station, DirectNet, or even a wireless adaptor for seamless wired to wireless networking.
- 12.2” TFT-LCD display.
- 6 standard channels, 8 optional channels.
- 24 hours of trended parameter data storage.
- Lithium-ion battery.
- Neonatal, pediatric, and adult monitoring capabilities.
Drager Infinity Delta XL Specifications
Dimensions
- Height: 10.7 in (27.2 cm)
- Width: 15.1 in (38.4 cm)
- Depth: 7.5 in (19 cm)
- Weight: 13.6 lbs (6.2 kg) (Without external battery)
Display
- Type: Thin Film Transistor-Liquid Crystal Display Active Matrix (TFT-LCD)
- Size (Diagonal): 12.2 in (31 cm)
- Resolution: 800 x 600
- Channels: 6 Standard, 8 Optional
- Viewing Area: 9.7” x 7.3” (24.6 x 18.5 cm)
Battery
- Type: Rechargeable Lithium-ion
- Capacity: 180 minutes
- Charging Time: 6.5 h
ECG
- Displays up to 12 leads
- Heart Rate Range: 15 to 300 bpm
- Accuracy: ±2 bpm or ±1% (whichever is greater)
Respiration
- Measuring Method: Impedance pneumography
- Measuring Range: 0 to 155 breaths per min
- Accuracy: ±1 breath/min or 2% of rate (whichever is greater)
- Apnea Detection: For neonatal and pediatric patients
SpO2
- Masimo SET, Nellcor OxiMax, Drager OxiSure
- Method: Transmission spectrophotometry
- SpO2 Measuring Range: 1 to 100%
- Pulse Measuring Range: 30 to 250 bpm
NBP (Noninvasive Blood Pressure)
- Displayed Parameters: Systolic, Mean and Diastolic pressures
- Measuring Method: Oscillometric utilizing step deflation
- Modes of Operation: Manual (single measurement); Continuous (5 minutes) and Interval
- Interval Times: 1, 2, 2.5, 3, 5, 10, 15, 20, 25, 30, 45, 60, 120 and 240 minutes
- Heart Rate Measuring Range: 30 to 240 bpm
IBP (Invasive Blood Pressure)
- Displays up to 8 pressures
- Measuring Method: Resistive strain gauge transducer
- Display Resolution: 1 mmHg
- Measuring Range: -50 to 400 mmHg (after zeroing)
- Frequency Ranges: DC to 8 Hz, DC to 16 Hz, or DC to 32 Hz (user-selectable)
- Zero Balance Range: ± 200 mmHg
Electrical
- Input Voltage: 11 to 15 V DC
- Power Consumption: ≤70 Watts (fully loaded)
- Patient Leakage Current: ≤10 mA
- Protection Class: Internally powered (per IEC 60601-1) and for use with specified Class 1 power supplies.
- Power Requirements: 100 to 240 V AC, 3 A
- Frequency: 50 to 60 Hz
Operating Specifications
- Temperature: 50° F to 104° F (10° C to 40° C)
- Humidity: 20% to 90%, non-condensing
- Atmospheric Pressure: 525 to 795 mmHg (70 to 106 kPa)
Storage Specifications
- Temperature: -4° F to 104° F (-20° C to 40° C)
- Humidity: 10% to 95%, non-condensing (With packaging)
- Atmospheric Pressure: 375 to 795 mmHg (50 to 106 kPa)