NIBP

Technique: Oscillometric (MAXNIBP® Technology)

Patient Range: Adult – Neonate

Measurement Range

Systolic: 30-255 mmHg

Diastolic: 15-220 mmHg

MAP: 20-235 mmHg

Pulse Rate Range: 30-240 BPM

Accuracy

Blood Pressure: ± 5 mmHg standard deviation no greater than 8 mmHg

Pulse Rate: ± 2% or ±2 BPM, whichever is greater

Oximetry (Options) Masimo SET®

SpO2 Range: 0-100%

SpO2 Accuracy:

70-100%, ± 2 digits (1 S.D.) Adult

70-100%, ± 3 digits (1 S.D.) Neonate Pulse Rate Range: 25-240 BPM Pulse Rate Accuracy: ± 3 BPM

Nonin®

SpO2 Range: 0-100%

SpO2 Accuracy:

70-100%, ± 2 digits (1 S.D.) Adult

70-100%, ± 3 digits (1 S.D.) Neonate

Pulse Rate Range: 18-240 BPM

Pulse Rate Accuracy: ± 3% or ±1 digit, whichever is greater

Nellcor® OxiMax® SpO2 Range: 1-100% SpO2 Accuracy:

70-100%, ± 2 digits (1 S.D.) Adult

70-100%, ± 3 digits (1 S.D.) Neonate Pulse Rate Range: 20-240 BPM Pulse Rate Accuracy: ± 3 digits

Temperature (Optional) Welch Allyn SureTemp® Plus

Temperature Range: 29 °C to 42 °C (84.2 °F to 107.6 °F)

Accuracy: ± 0.1 °C (± 0.2 °F) Meets or exceeds ASTM Standards

Control Panel

Display: LED display of measurement results, instructions, troubleshooting messages, and signal strength bar

Parameters Displayed: Systolic Pressure, Diastolic Pressure, Mean Arterial Pressure (MAP), Pulse Rate,

%SpO2, Temperature (°C or °F)

Patient Alarms

Adjustable high and low alarms for Pulse Rate, SpO2, Systolic Pressure, Diastolic Pressure and Mean Arterial Pressure

Operating Modes

History: Review of previous measurements

NIBP: Manual, STAT or Automatic (at preset intervals)

%SpO2Continuous Monitoring

Temperature: Predictive or Continuous

Power

Source: External line or internal battery

AC Power: 100-240 VAC, 50-60 Hz, 0.5A;

T1.25A L250V fuse rating

DC Power (EMS Option): 12-30 VDC Battery: NiMH battery pack (removable) Charge Time: 4 hours

Operating Time: Minimum 4 hours of continuous operation

Leakage Current: 100 µA (maximum)

Safety Limits

145 mmHg (Neonate)

If measurement time exceeds 120 seconds (Adults), 90 seconds (Neonate)

If safety timer detects microprocessor failure

Operating Environment

Operating Temperature: 0° C to 50° C (32° F to 122° F)

Storage Temperature: -20° C to 65° C (-4° F to 149° F)

Humidity: 15 – 95%, non-condensing

Altitude: Up to 3048 meters (10,000 feet)

if stored or used outside temperature and humidity ranges.

Physical Dimensions & Weight (Base Unit)

H x W x D: 17.0 cm x 21.5 cm x 7.5 cm (6.75 in x 8.5 in x 3.0 in)

Weight: 1.4 kgs (3 lbs)

Standards

ANSI/AAMI SP-10

EN 60601-1, EN 60601-1-2, EN 865, EN 60601-2-30,

EN 60601-2-49

ETL Listed – UL 60601, CAN/CSA C22.2 No. 601.1

CE Marking according to Directive 93/42/EEC

Optional Accessories

Infrared printer RS-232 interface

Swiveled hard mount (for ambulance applications) Roll stand and basket